Specification
- Conductivity cell covers wide range µS to mS.
- Stable & quick response.
- Intended for laboratory use.
- Glass-body conductivity cells are chemical-resistant. They can be used in aqueous samples as well as in samples containing organic solvents.
- Typically, a low conductivity solution is measured with a low cell having a k=0.1 cm-1 cell constant to prevent absorption of carbon dioxide (CO₂) from air, which can affect the conductivity value.
- A high conductivity solution is measured with a high cell having a k=1.0 cm-1 cell constant.
- Long life, easy cleaning to avoid contamination of the electrode to ensure correct results.
- The same electrodes can also be used for measuring Resistivity / TDS / Salinity.
- Available with BNC, 5 pin connectors & Banana Pin.
- Suitable/ compatible with mostly all make Meters/ Analysers.